Intended Use
Recipient: 1 / Purple / K2-EDTA / 2 mL
- Sample collection
- Recipient: Fasting Blood Samples preferred
- Specimen preparation: DO NOT FREEZE blood or tissue samples intended for HLA testing or crossmatching, as freezing damages the cells.
- Storage/Transport Temperature: Samples may be stored temporarily at refrigerated temperatures (4°C to 8°C) prior to dispatch. Ensure they are packaged to maintain this temperature during transit if required by local protocols.
- Unacceptable Conditions
- Sample older than 24 hours after collection is not suitable.
- Lipemic samples will interfere with the reading of results.
- Haemolysed Samples.
- Sample collected within 24 hours of dialysis will not be accepted.
- Remarks –
- Stability –
Test performed using the One Lambda LABScreenTM Single Antigen assays for detection of HLA Class I and HLA Class II antibodies in the patient’s serum/ plasma. Test serum is incubated with LABScreen beads. Any HLA antibodies present in the test serum bind to the antigens on the beads and then are labeled with R-Phycoerythrin (PE)- conjugated goat anti-human IgG. The Luminex Labscan 3D flow analyzer simultaneously detects the fluorescent emission of PE and a dye signature from each bead, allowing almost real-time data acquisition. To assign PRA and HLAspecificity, the reaction pattern of the test serum is compared to the lot-specific worksheet defining the antigen array.
Sample received to report Turnaround time (TAT)
1 Working Day
————-
Interpretive Data
————–
Resources
- Additional Technical Information
- Test Request Form
- List Item
Sample Reports
- Enhanced Report
- See report